GUI reference
R&S
®
ZNA
753User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate<Chn>:EYE:MASK:STATe
CALCulate<Chn>:EYE:MASK:DATA?
CALCulate<Chn>:EYE:MASK:FAIL?
Mask Fail Beep
This checkbox determines whether the R&S ZNA should make an audible beep on
mask failures.
Note: The R&S ZNA does not have a built-in audio device and loudspeaker. To hear
these sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
CALCulate<Chn>:EYE:MASK:FAIL:BEEP
Export Test Results
Opens a dialog that allows the user to save the mask test results to an ASCII file.
Remote command:
MMEMory:STORe:EYE:MASK:RESults
Global Check, TTL1 Pass, TTL2 Pass
see Chapter 5.6.1, "Limit Test tab", on page 431