Performing measurements
R&S
®
ZNA
65Getting Started 1178.6456.02 ─ 18
Due to the R&S ZNA's calibration wizard, calibration is a straightforward, guided proc-
ess.
1. Replace the DUT by the Through standard of your calibration kit. Make sure to dis-
connect all calibration units.
2. Select Channel – [Cal] > "Configure/Start Calibration..." to open the calibration
setup dialog.
3. "Delete" all except the first row in the "Cal Settings" table. If the table is empty, use
"Add" to create a calibration setup.
4. Tap on the "Calibration" cell of the first row in the "Cal Settings" table to open the
"Define Calibration" dialog.
5. Select port 1 (P1) and port 2 (P2) and the calibration type "Trans Norm".
6. Select "OK" to apply your settings and return to the calibration setup dialog "Cali-
bration Setting" wizard.
7. Select the test port connector type and gender (here: N 50 Ω, female, correspond-
ing to a male Through standard), and the calibration kit (here: R&S ZV-Z121).
Transmission S-parameter measurement