Command reference
R&S
®
ZNA
1537User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]FREQuency:MDELay:CONVersion <Mixer delay conversion flag>
Disables the mixer delay measurement and switches back to normal (non frequency-
converting) mode.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Mixer delay
conversion flag>
OFF
Disable mixer delay measurement
*RST: OFF
Example: See [SENSe<Ch>:]FREQuency:MDELay:APERture
on page 1535
Usage: Setting only
Manual operation: See "Apply/OK" on page 373
[SENSe<Ch>:]FREQuency:MDELay:CORRection[:STATe]
<arg0>
Qualifies whether the analyzer applies the current mixer delay measurement calibra-
tion data.
Tip:
The analyzer automatically applies the calibration data when a calibration file is loaded
(MMEMory:LOAD:MDCData) or after a calibration sweep has been completed
([SENSe<Ch>:]FREQuency:MDELay:ACQuire).
Suffix:
<Ch>
.
Channel number.
Parameters:
<arg0> ON | OFF – Apply / ignore calibration data
*RST: OFF
Example: See MMEMory:LOAD:MDCData on page 1539
Manual operation: See "Mixer Dly Cal Active" on page 372
[SENSe<Ch>:]FREQuency:MDELay:DIVide <Group Delay Divide by 2 flag>
Suffix:
<Ch>
.
Parameters:
<Group Delay Divide
by 2 flag>
SCPI command reference