EasyManua.ls Logo

R&S ZNA67 - Offset Tab

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
661User Manual 1178.6462.02 ─ 20
Active
The checkboxes in the "Active" column activate or deactivate the configured differential
match embedding (i.e. adds or removes it to/from the calculation flow) without chang-
ing its parameters.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:
EMBedding<LogPt>[:STATe]
File Name 1
The ellipsis button in the "File Name 1" column is enabled if a 2-port data Network is
selected.
When loading the touchstone file (*.s2p), the analyzer by default assumes odd ports
left (VNA side), even ports right (DUT side). However, it is also possible to "Swap
Gates", instructing the analyzer to reinterpret the loaded S-parameters (e.g. S
12
-->
S
21
).
Note: The loaded S-parameter traces are stored in the active recall set. Persisted
recall sets contain the full embedding and deembedding data so that they can be trans-
ferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:DIFFerential:EMBedding<LogPt>
5.14.2 Offset tab
Defines length offset parameters for each port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see Chapter 5.14.1.2, "Offset panel", on page 652).
The marker function Zero Delay at Marker function overwrites the offset parameters.
Background information
Refer to the following sections.
Chapter 4.6, "Offset parameters and de-/embedding", on page 201
Chapter 4.6.1.1, "Definition of offset parameters", on page 201
Chapter 4.6.1.3, "Auto Length", on page 202
Chapter 4.6.1.6, "Application and effect of offset parameters", on page 206
Chapter 4.6.1.7, "Offset parameters for balanced ports", on page 207
Chapter 4.6.1.5, "Fixture Compensation", on page 205
Offset Embed softtool