687User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate:FMODel:EZD<Ph_pt>:GENerate:SIDE<1|2>
DC Extrapolation
If unchecked (default), the touchstone data of the generated deembedding file reach
down to the start frequency of the instrument sweep. If checked, the EZD tool adds
(extrapolated) S-parameter data for f = 0.
Remote command:
CALCulate:FMODel:EZD<Ph_pt>:DCEXtrapolat
Impedance Corrected/Impedance Correction Settings
Disables/enables impedance correction and provides related settings.
Remote command:
CALCulate:FMODel:EZD<Ph_pt>:IMPedance on page 942
CALCulate:FMODel:EZD<Ph_pt>:IMPedance:REFerence on page 943
CALCulate:FMODel:EZD<Ph_pt>:IMPedance:BWIDlimit on page 942
5.14.6 Port Sets tab
Selects transformation networks for deembedding/embedding arbitrary port sets,
defines their parameters, assigns them to a port set and enables embedding.
Use the complementary dock widget to create the required port sets and to activate or
deactivate dembedding/embedding for selected port sets (see Chapter 5.14.1.4, "Port
Sets panel", on page 655).
Background information
Refer to Chapter 4.6.2.5, "Port pair de-/embedding", on page 211 and Chapter 4.6.2.6,
"Port set de-/embedding", on page 212.