GUI reference
R&S
®
ZNA
787User Manual 1178.6462.02 ─ 20
● Presets tab............................................................................................................787
● Calibration tab.......................................................................................................788
● User Interface tab..................................................................................................791
● Messages tab........................................................................................................793
● Channel Bits tab....................................................................................................794
● Advanced tab........................................................................................................ 795
● Power tab..............................................................................................................797
● Recovery tab.........................................................................................................800
● HUMS tab..............................................................................................................801
Presets tab
Specifies the behavior of the R&S ZNA upon a preset.
Preset Scope
Defines whether a preset affects all open recall sets ("Instrument") or the active recall
set only.
The "Preset Scope" applies to the GUI and to the SYSTem:PRESet[:DUMMy] com-
mand. The *RST command always resets all open recall sets.
Remote command:
SYSTem:PRESet:SCOPe
Remote Preset Configuration
"Align *RST to User Defined Preset" defines the behavior of the *RST and
SYSTem:PRESet commands. If checked, and if a valid setup file is selected in the Pre-
set Configuration section, then they perform a user-defined preset. Otherwise, they
perform a factory preset.
Note that the preset modes "RF Off" and "Normal, GUI, Ext Setup" are not available via
remote commands.
Remote command:
SYSTem:PRESet:REMote[:STATe]
Setup softtool