Command reference
R&S
®
ZNA
885User Manual 1178.6462.02 ─ 20
Use CALCulate:CALValidate:STANdard to select the CalU standard to be used
for the validation.
Parameters:
<String> *RST: 'Factory'
Manual operation: See "Characterization" on page 557
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer <Port>, <Value>
Defines the maximum magnitude deviations from the reference characterization S-
parameters that are allowed during cal validation.
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.
Parameters:
<Port> Physical analyzer port no.
<Value> *RST: -0.2 (LOWer) | +0.2 (UPPer)
Default unit: dB
Manual operation: See "Validation Limits" on page 557
CALCulate:CALValidate:RESPonse:PHASe:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:PHASe:UPPer <Port>, <Value>
Defines the maximum phase deviations from the reference characterization S-parame-
ters that are allowed during cal validation.
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.
Parameters:
<Port>
<Value> *RST: -2 (LOWer) | +2 (UPPer)
Default unit: degrees
Manual operation: See "Validation Limits" on page 557
CALCulate:CALValidate:RUN
Runs the cal validation with its configured settings.
Usage:
Event
Manual operation: See "Validate" on page 556
CALCulate:CALValidate:STANdard <Standard>
Selects the CalU standard to be used for cal validation.
SCPI command reference