Command reference
R&S
®
ZNA
957User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:SFD<Ph_pt>:DIFFcfg <SFDPortConfig>
Tells the SFD tool about the port ordering of the 2x Thru test coupon.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<SFDPortConfig> ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
Manual operation: See "2x Thru Port Ordering" on page 685
CALCulate:FMODel:SFD<Ph_pt>:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure the DUT +
Fixture at the active ports (see CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]).
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Measure" on page 678
CALCulate:FMODel:SFD<Ph_pt>:DUT:MEASure:FILename <String>
For a fixture modeling with the EZD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Manual operation: See "Load File" on page 678
CALCulate:FMODel:SFD<Ph_pt>:DUT:MEASure:FILename:CLEar
Invalidates the file path previously set using CALCulate:FMODel:SFD<Ph_pt>:
DUT:MEASure:FILename.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Load File" on page 678