Command reference
R&S
®
ZNA
983User Manual 1178.6462.02 ─ 20
CALCulate<Chn>:MARKer<Mk>:TYPE.......................................................................... 1005
CALCulate<Chn>:MARKer<Mk>:X.................................................................................1006
CALCulate<Chn>:MARKer<Mk>:Y.................................................................................1006
CALCulate:MARKer:COUPled[:STATe] <Boolean>
Enables marker coupling to the active trace of the active channel or disables it.
Use CALCulate:MARKer:COUPled:TYPE to select the suitable coupling type before
setting CALCulate:MARKer:COUPled:STATe to ON.
Parameters:
<Boolean> ON | OFF - enables or disables marker coupling.
*RST: OFF
Example: Suppose that the active recall set contains traces Trc1 and Trc2,
assigned to channels no. 1 and 2, respectively.
:CALC2:PAR:SEL 'TRC2'; :CALC2:MARK1 ON; MARK2
ON
Select Trc2 as the active trace of channel 2 and activate mark-
ers 1 and 2 for it. The default position for both markers is the
center of the sweep range.
INSTrument:NSELect 2
Make channel 2 the active channel.
CALCulate:MARKer:COUPled:TYPE ALL
Select coupling type "All" (recall set-wide coupling).
CALC:MARK:COUP ON
Activate marker coupling to Trc2 (the active trace of the active
channel). Automatically creates markers no. 1 and 2 on Trc1
and couples them to the respective markers of Trc2.
Manual operation:
See "Coupled Markers" on page 453
CALCulate:MARKer:COUPled:TYPE <Type>
Selects the marker coupling type to be used in the current recall set.
Marker coupling is enabled using CALCulate:MARKer:COUPled[:STATe] ON.
Changing the coupling type automatically disables marker coupling throughout the
recall set.
Parameters:
<Type> ALL | CHANnel | DIAGram
Marker coupling mode
ALL
recall set-wide
CHANnel
channel-wide
DIAGram
diagram-wide
SCPI command reference