EasyManua.ls Logo

R&S ZNA67 - Calculate<Chn>:Marker<Mk>:Function:start; Calculate<Chn>:Marker<Mk>:Function:stop; Calculate<Chn>:Marker<Mk>:Mode

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
996User Manual 1178.6462.02 ─ 20
Example:
*RST; :CALC:MARK ON; MARK:DELTa ON
Create marker 1 in the center of the current sweep range and
enable the delta mode.
CALC:MARK:X 300MHz
Increase the stimulus value of the delta marker by 300 MHz.
CALC:MARK:FUNC:SPAN
Set the sweep range equal to 300 MHz. The sweep range starts
at the reference marker position, i.e. in the center of the analyz-
er's frequency range.
Usage: Event
Manual operation: See "Center = Marker / Start = Marker / Stop = Marker / Span =
Marker" on page 470
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP
These command sets the beginning (...STARt) and the end (...STOP) of the sweep
range equal to the stimulus value of the marker <Mk> on trace no. <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number.
Example:
*RST; :CALC:MARK ON
Create marker 1 in the center of the current sweep range and
assign it to trace no. 1.
CALC:MARK:FUNC:STAR
Divide the sweep range in half, starting at the current marker
position. As an alternative:
CALC:MARK:FUNC:STOP
Divide the sweep range in half, ending at the current marker
position.
Usage: Event
Manual operation: See "Center = Marker / Start = Marker / Stop = Marker / Span =
Marker" on page 470
CALCulate<Chn>:MARKer<Mk>:MODE <Mode>
Sets marker no. <Mk> to continuous or discrete mode. The marker doesn't have to be
created before (CALCulate<Chn>:MARKer<Mk>[:STATe] ON), the mode can be
assigned in advance.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number.
SCPI command reference

Table of Contents

Other manuals for R&S ZNA67