Alignment R&S ZVA
1177.5333.02 2.4 E-11
Aligning the Frequency Accuracy
Spectrum analyzer (section "Test Equipment", item 1):
Error <1x10
-9
Connect the spectrum analyzer to the 10-MHz reference output at the
rear of the ZVA.
Select internal reference
MENU : System: Reference Internal
Spectrum analyzer settings:
Center frequency: 10 MHz
Span: 200 Hz
Before the following measurement is performed, the ZVA must have
been switched on for at least 30 minutes to give the reference
oscillator time to warm up.
Measure the frequency with the spectrum analyzer:
Nominal frequency:
Model without OCXO (Option B4) 10 MHz ± 80 Hz
Model with OCXO (Option B4) 10 MHz ± 1 Hz
Alignment without Option ZVAB-B4:
Enter Service Function 2.1.1.6.209.0x000000
Read off the frequency, e.g. 10.000050 MHz.
Change the right-hand segment (corresponding to bit 0 to bit 11) of the data word - e.g. to
000400 - instead of 000000.
Read off the frequency again, e.g. 10.000010 MHz.
Change the left-hand segment of the data word, until the counter indicates precisely
10.000000 MHz.
Alignment with Option ZVAB-B4:
Enter Service Function 2.1.1.6.209.0x800000.
Read off the frequency, e.g. 10.000005 MHz.
Change the left-hand segment (corresponding to bit 12 to bit 23) of the data word - e.g. to
400000 - instead of 800000.
Read off the frequency again, e.g. 10.000001 MHz.
Change the left-hand segment of the data word, until the counter indicates precisely
10.000000 MHz.