R&S ZVA Repairs
1177.5333.02 3.115 E-11
Testing the Reflectometer RM67
Generator levels
It is assumed that the synthesizer section (synthesizer1) associated with the reflectometer to be tested
is OK.
Loosen cable W514A (GEN -> BIAS 67) or W512A (GEN -> attenuator) at both ends and screw off
at the generator output GEN.
Connect the generator output to the spectrum analyzer using the V-cable (if FSU50 or FSU67 is not
available use a power meter for level measurements).
Set the ZVA to the CW sweep mode.
Set the frequencies listed in the table.
Set the power listed in the table.
Set power to
(without GenAtt)
Set power to
(with GenAtt)
With the stated levels, the cable loss must still be taken into account.
Receiver levels
The following is assumed:
The LO synthesizer section associated with the reflectometer to be tested is OK.
The network controller associated with the reflectometer to be tested is OK.
One reflectometer in the instrument is functioning.
Loosen the cable to the receiver input MEAS (X6) and the cable to the receiver input REF (X5) at
both ends and screw off at the receiver inputs.