EasyManua.ls Logo

React Health PO2BLK - User Manual

React Health PO2BLK
2 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Sarasota, FL 34243
Info@ReactHealth.com
www.Reacthealth.com
PO2BLK
PO2BLU
PULSE OXIMETER USER MANUAL
Instructions to User
The information contained in this document is subject to change without notice.
The Manual describes, in accordance with the Pulse Oximeter’s features and requirements, main structure, functions, specications, correct methods for
transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and equipment.
Refer to the respective chapters for details.
Please read the User Manual carefully before using this product. Follow directions carefully for proper use. Manufacturer is not responsible for any issues
caused by improper use of the device. This product can be used repeatedly and has an estimated operating life of 3 years. If you have any questions or
problems with use of this device, please visit us at www.reacthealth.com and navigate to our Support tab for contact options, including submission of an
online support ticket.
WARNING:
Uncomfortable or painful feeling may develop with continuous use. It is recommended that the sensor should not be applied to the same finger for
over 2 hours.
The device can not be clipped on fingers with edema or tender tissue.
The light (the infrared is invisible) emitted from the device is harmful to the eyes, so keep away from eyes.
User cannot use nail polish or artificial nails. As it may cause false readings.
Long nails may impede proper readings.
Please refer to the correlative literature about the clinical restrictions and cautions.
This device is not intended for treatment.
The User Manual is published by React Health. All rights reserved.
1 Safety
1.1 Instructions for Safe Operations
Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect user’s safety and monitoring
performance including cables and transducers. It is recommended that the device should be inspected once a week. When there is obvious damage, stop
using the oximeter .
Necessary maintenance must be performed by qualied engineers ONLY. Users are not permitted to service it by themselves.
The oximeter cannot be used together with devices not specied in User’s Manual. Only the accessories recommended by manufacture can be used
with this device.
This product is calibrated before leaving factory.
1.2 Warnings
Explosive hazard—DO NOT use the oximeter in environment with inammable gas such as some ignitable anesthetic agents.
This device contains rubber.
The disposal of scrap instrument and its accessories and packings (including battery, plastic bags, foams and paper boxes) should follow the local laws
and regulations.
Please check the packing before use to make sure the device and accessories are in accordance with the packing list.
Do not service while patient is wearing the pulse oximeter.
No modication of this equipment is allowed.
The user is an intended operator.
The probe of the device is the applied part.
1.3 Attentions
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
If the oximeter gets wet, please do not use.
Allow pulse oximeter to reach room temperature if it has been stored n a hot or cold environment.
DO NOT operate keys on front panel with sharp materials.
High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to User Manual in the relative chapter for instructions on
cleaning and disinfection.
Do not immerse oximeter in water or other cleaning uids. Device may be cleaned with an alcohol wipe and a soft cloth.
When cleaning the device with water, the temperature should be lower than 60˚C.
Very thin or cold ngers may not provide accurate readings.
Do not use the device on infant or neonatal users.
The product is suitable for adults over 40kg.
The device may not work for all users. If you are unable to achieve stable readings, discontinue use.
The update period of data is less than 5 seconds, which is changeable according to different individual pulse rate.
If some abnormal conditions appear on the screen during test process, pull out the nger and reinsert to restore normal use.
The lanyard attached the product is made from hypo-allergenic material, if particular group are sensitive to the lanyard, stop using it. In
addition, pay attention to the use of the hanging rope , do not wear it around the neck to avoid the risk of strangulation.
The instrument dose not have low-voltage alarm function, it only shows the low-voltage, please change the battery when the icon shows on the screen.
This pulse oximeter does not have audible alarms.
Batteries must be removed if the device is going to be stored for more than one month.
A exible circuit connects the two parts of the device. Do not twist or pull on the connection.
1.4. Indication for Use
The Pulse Oximeter is a non-invasive device intended for the spot-check of saturation of arterial hemoglobin(SpO
2
) and the pulse rate of adult in home use
environments. This device is not intended for continuous monitoring. The device can be multi-used. Intended to monitor heart rate during exercise.
2 Overview
The pulse oxygen saturation is the percentage of HbO
2
in the total Hb in the blood, the O
2
concentration in the blood. It is an important bio-parameter to
monitor for respiratory function. This device will simultaneously measure the oxygen saturation of the blood and the pulse rate.
The Pulse Oximeter features low power consumption, portability and ease of use. Diagnosis of low SpO
2
is as simple as placing the device on your nger.
2.1 Features
Operation of the product is simple and convenient.
The product is small and lightweight.
Power consumption of the product is low and the two originally equipped AAA batteries can be operated continuously for 24 hours.
The product will automatically be powered off when no signal, within 5 seconds.
Low-battery indicator displays as a ashing batery indicator.
2.2 Major Applications and Scope of Application
The Pulse Oximeter can be used in measuring pulse oxygen saturation and pulse rate through nger. The product is suitable for family use(It can be used before
or after doing sports, and it is not recommended to use the device during the process of doing sports).
! This product is not intended to diagnose or treat carbon monoxide patients.
2.3 Environment Requirements
Storage Environment
a) Temperature: -40˚C~+60˚C
b) Relative humidity: ≤95%
c) Atmospheric pressure: 500hPa~1060hPa
Operating Environment
a) Temperature: 10˚C~40˚C
b) Relative Humidity: ≤75%
c) Atmospheric pressure: 700hPa~1060hPa
3 Principle and Caution
3.1 Principle of Measurement
Principle of the Oximeter is as follows: An experience formula of data process is established taking use of Lambert Beer Law according to Spectrum
Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin (HbO
2
) in glow & near-infrared zones. Operation principle of the instrument
is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning & Recording Technology, so that two beams
of different wavelength of lights can be focused onto human nail tip through perspective clamp nger-type sensor. Then measured signal can be obtained by a
photosensitive element, information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor.
3.2 Caution
1. The nger should be placed properly (see the attached illustration of this manual, Figure 7), or else it may cause inaccurate measurement.
2. The SpO
2
sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in a position there between.
3. The SpO
2
sensor should not be used at a location or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection.
4. Make sure the optical path is free from any optical obstacles like fabric or articial nails.
5. Excessive ambient light may affect the measuring result. This includes uorescent lamp, dual ruby light, infrared heater, direct sunlight and etc.
6. Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
7. User can not use enamel or articial nails.
3.3 Clinical Restrictions
1. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood ow of subject is required. For a subject with weak pulse due to
shock, low ambient/body temperature, major bleeding, or use of vascular contracting drug, the SpO
2
waveform (PLETH) will decrease. In this case, the
measurement will be more sensitive to interference.
2. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or carbon monoxide
hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, the SpO
2
determination by this monitor may be
inaccurate.
3. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for serious error of SpO
2
measure.
4. As the SpO
2
value serves as a reference value for judgement of anemic anoxia and toxic anoxia, some users with serious anemia may also report good
SpO
2
measurement.
4 Technical Specifications
1) Display Format: Digital tube Display;
SpO
2
Measuring Range: 0% - 100%;
Pulse Rate Measuring Range: 30 bpm - 250 bpm;
Pulse Intensity Display: columniation display
2) Power Requirements: 2 ×1.5V AAA alkaline battery, adaptable range: 2.6V-3.6V.
3) Power Consumption: Smaller than 25 mA.
4) Resolution: 1% for SpO
2
and 1 bpm for Pulse Rate.
5) Measurement Accuracy: ±2% in stage of 70%-100% SpO
2
, and meaningless when stage being smaller than 70%. ±2 bpm or±2% (select larger) for
Pulse Rate. Clinical Trial :SpO
2
regression plot & Bland–Altman plot,Refer to Figure 2 & Figure 3.
6) Measurement Performance in Weak Filling Condition: SpO
2
and pulse rate can be shown correctly when pulse-lling ratio is 0.4%. SpO
2
error is
±4%, pulse rate error is ±2 bpm or ±2% (select larger).
7) Resistance to surrounding light: The deviation between the value measured in the condition of man-made light, indoor natural light and that of
darkroom is less than ±1%.
8) It is equipped with a switch function. The Oximeter can be powered off when the nger is off the oximeter within 5 seconds.
9) Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
5 Accessories
Lanyard
Carry Case
Two batteries
Silicone Cover
6 Installation
6.1 View of the Front Panel
Figure 2 SpO
2
regression plot
Figure 3 Bland–Altman plot
Figure 4. Front View Figure 5. Batteries Installation
6.2 Battery
Step 1. Refer to Figure 5. and insert the two AAA size batteries properly in the right direction.
Step 2. Replace the battery cover. Please take care when you insert the batteries for the improper insertion may damage the device.
! Please take care when you insert the batteries for the improper insertion may damage the device.
6.3 Mounting the Lanyard
Step 1. Put the end of the rope through the hole.
Step 2. Put another end of the rope through the rst one and then tighten it.
Figure 6. Mounting the lanyard Figure 7. Put nger in position
7 Operating Guide
7.1 Insert the two batteries and replace the cover.
7.2 Open the clip as shown in Figure 7.
7.3 Insert the user’s nger between the rubber cushions (make sure the nger is in the right position), and then clip the nger.
7.4 Press the switch button once on front panel.
7.5 Ensure the user is still, without excessive hand movements.
7.6 The users pulse rate and SpO
2
will display on screen.
7.7 Press the button to reset the device.
! Fingernails and the luminescent tube should be on the same side.
8 Repairing and Maintenance
Please change the batteries when the low-voltage displayed on the screen.
Surface clean your device with an alcohol wipe and a soft cloth before using.
To avoid cross contamination, wipe with alcohol between use.
Please remove the batteries if the oximeter is not in use for a long time.
The packed device can be transported by ordinary conveyance or according to transport contract. The device can not be transported mixed with toxic,
harmful, corrosive material.
The best storage environment of the device is - 40ºC to 60ºC ambient temperature and not higher than 95% relative humidity.
! High-pressure sterilization cannot be used on the device.
Figure 1. Operating Principle
Question and Answer IconNeed help?

Do you have a question about the React Health PO2BLK and is the answer not in the manual?

Overview

The PO2BLK PO2BLU device is a pulse oximeter designed for home use. It measures oxygen saturation (SpO2) and pulse rate, providing users with a simple and convenient way to monitor these vital signs. The device is intended for non-invasive spot-check measurements of functional oxygen saturation of arterial hemoglobin and pulse rate in adults.

Function Description

The pulse oximeter operates by emitting two different wavelengths of light (red and infrared) through a finger. A photodetector on the opposite side of the finger measures the amount of light absorbed. Since oxygenated and deoxygenated hemoglobin absorb light differently, the device can calculate the percentage of oxygenated hemoglobin in the blood, which is the SpO2. Simultaneously, the pulsatile nature of arterial blood flow allows the device to determine the pulse rate.

The device features a bright OLED display that shows the SpO2 value, pulse rate, pulse bar, and a plethysmograph waveform. The plethysmograph provides a visual representation of the blood flow and can help assess the signal quality. The display has adjustable brightness and can be rotated in four directions for easy viewing.

The device is designed for ease of use, with a simple one-button operation. It automatically powers off after 8 seconds of inactivity to conserve battery life. It also includes a low battery indicator to alert the user when the batteries need replacement.

Important Technical Specifications

Display Type: OLED display SpO2 Measurement Range: 70% - 100% SpO2 Measurement Accuracy: ±2% (70%-100%), unspecified below 70% SpO2 Resolution: 1% Pulse Rate Measurement Range: 30 bpm - 250 bpm Pulse Rate Measurement Accuracy: ±2 bpm or ±2% (whichever is greater) Pulse Rate Resolution: 1 bpm Perfusion Index (PI) Range: 0.2% - 20% Power Requirements: Two 1.5V AAA alkaline batteries Power Consumption: Less than 30mA Battery Life: Approximately 24 hours of continuous operation Dimensions: 58mm (L) × 31mm (W) × 31mm (H) Weight: Approximately 50g (including batteries) Operating Temperature: 5°C - 40°C Storage Temperature: -10°C - 40°C Operating Humidity: 15% - 80% (non-condensing) Storage Humidity: 10% - 80% (non-condensing) Atmospheric Pressure: 700hPa - 1060hPa Storage Atmospheric Pressure: 500hPa - 1060hPa Ingress Protection: IP22 (protected against solid objects greater than 12.5mm and vertically falling water drops when tilted up to 15°) Safety Classification: Type BF applied part

The device is designed to meet EMC standards for medical electrical equipment, specifically IEC 60601-1-2. It is classified as Group 1, Class B for electromagnetic emissions, meaning it is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Usage Features

Preparation for Use:

  1. Install two AAA batteries into the battery compartment, ensuring correct polarity.
  2. Open the clamp of the pulse oximeter.
  3. Insert one finger into the rubber opening of the oximeter, ensuring the nail is facing upwards.
  4. Release the clamp.
  5. Press the power button once on the front panel.
  6. The measurement results will be displayed on the screen.

Measurement Guidelines:

  • The finger should be clean and free of nail polish, artificial nails, or any other obstructions that might interfere with light transmission.
  • Avoid excessive movement during measurement.
  • The device is intended for spot-check measurements, not continuous monitoring.
  • If the reading is unstable, try repositioning the finger or using a different finger.
  • The device should not be used on the same limb simultaneously with a blood pressure cuff or arterial duct infusion.
  • The device is not intended for use on infants or neonates.
  • The device should not be used in an MRI or CT environment.
  • Avoid strong ambient light, as it may affect the accuracy of the measurement.
  • If the reading is too low or too high, consult a medical professional.

Display Features:

  • SpO2 Value: Displays the oxygen saturation percentage.
  • Pulse Rate Value: Displays the heart rate in beats per minute.
  • Pulse Bar: A bar graph that indicates the pulse intensity.
  • Plethysmograph Waveform: A real-time graph of the blood flow, providing a visual indication of signal quality.
  • Low Battery Indicator: An icon that appears when the batteries are low.
  • Display Direction: The display can be rotated in four directions by pressing the power button briefly during measurement.
  • Brightness Adjustment: The display brightness can be adjusted by holding the power button for more than one second while the device is on.

Maintenance Features

Cleaning:

  • Clean the surface of the device with a soft cloth dampened with 75% alcohol.
  • Do not immerse the device in water or other cleaning solutions.
  • Ensure the device is dry before reuse.
  • Clean the inside of the finger clip with a soft cloth dampened with 75% alcohol between uses.

Disinfection:

  • The device can be disinfected using 75% alcohol.
  • Ensure the device is dry before reuse.

Battery Replacement:

  • When the low battery indicator appears, replace the two AAA batteries.
  • Ensure correct polarity when inserting new batteries.
  • Remove batteries if the device will not be used for an extended period to prevent leakage.

Storage:

  • Store the device in a cool, dry place, away from direct sunlight, extreme temperatures, and humidity.
  • Keep the device out of reach of children.

Troubleshooting:

  • No SpO2/Pulse Rate displayed: Check battery installation, replace batteries, ensure finger is inserted correctly, and ensure the device is not damaged.
  • Unstable readings: Ensure the finger is still, clean, and free of obstructions. Try a different finger.
  • Power off automatically: This is a normal feature to conserve battery life. If it powers off too quickly, check battery level.
  • Display too dim: Adjust the display brightness.

The device has an expected service life of five years. Regular inspection for damage or wear is recommended. If any damage is found, contact the manufacturer or service center for repair. Do not attempt to repair the device yourself.

React Health PO2BLK Specifications

General IconGeneral
BrandReact Health
ModelPO2BLK
CategoryMedical Equipment
LanguageEnglish