TROUBLESHOOTING
1. THE DEVICE DOES NOT TURN ON.
lf the device does not turn on, please check CA) whether
the power plug has been correctly plugged into the socket
and/or whether the device is turned on. lf the unit is still
not turning on, unplug the power plug (B) and unscrew the
fuse holder with the inscription “FUSE” on the bottom of
the unit (see picture on the left) completely (with a
screwdriver or a small coin) and check that the fuse inside
is intact. (see picture on the right).
The fuse holder is located
on the
bottom of the device.
lntact fuse: the wire in the
fuse is undamaged.
(C) lf the fuse is intact, screw the fuse holder with the fuse
back into the device completely. lf the fuse is damaged,
replace the fuse with a commercially-available fine fuse
(TR); 3.15 A; 5 x 20 mm). You are also welcome to contact
our customer service. (D) Note that the device may only be
powered by 220 V power.
2. THE DEVICE IS PUMPING, BUTT ONLY BLOWS ONE OR
NONE OF THE TWO CUFFS.
lf only one or neither of the two cuffs should inflate after
starting the application, please check (A) whether the multi-
plug is fully and correctly inserted into the device. Also, make
sure (B) that the air hoses are not kinked, for example, because
you are sitting on the hoses. Check (C) that the gray cuff plugs
are all correctly attached to the cuff.
38