Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device sho
uld
assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment
- guidance
Electrostatic
discharge (ESD) IEC
61000
-4-2
±
8 kV contact
±
15 kV air
±
8 kV
contact
±15 kV air
Floors should be wood, concrete or ceramic tile. lf floor are
covered with synthetic materi
- al, the relative humidity
should be at least 30%.
Electrical fast transient/burst IEC
61000
-4-4
±
2 kV for power supply lines
±
2kV for power supply
lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge IEC 61000
-4-5
±
1 kV line(s) to line(s)
±
1 kV
differential mode
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short interruptions and
voltage variations on power supply
input lines IEC 61000
-4-11
<5% UT (>95% dip
in UT) for 0.5 cycle 40% UT (60% dip
in UT) for 5 cycles 70%UT(30%
dip in UT) for 25 cycles <5% UT (>95%
dip in UT) for 5 sec
<5% UT (>95%
dip in UT) for
0.5 cycle 40% UT (60% dip
in UT) for 5 cycles 70% UT
(30% dip in UT) for25
cycles <5% UT (>950/odip
in UT) for 5 sec
Mains power quality should be that of a typical commercial
or hospital environment. lf the user of the DEVICE requires
continued operation during power mains interruptions,
it is recommended that the DEVICE be powered from an
uninterruptible power supply or a battery.
Power frequency (50Hz/60Hz)
magnetic field IEC 61000
-4-8
l0V/m
l0V/m
Power frequency magnetic field should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Note: Ur is the AC mains voltage prior to application of the test level.
46