Operating Instructions
Insertthemicrophoneundertestingintothecalibrator’smicrophone
cavityandensurethatitisfullyinserted(soundlevelmeterswith1/2”
microphonescanbecalibratedwiththisdevice).
Slidetherangeswitchtothe94dBposition,orthe114dBpositionas
desired.Thecalibratorgeneratesasignalassoonastheswitchisset.
Readthesoundlevelmeter’svalue.Thesoundlevelmetershouldread
94dBor114dB,dependingonthecalibrator’ssignallevel.Ifthesound
levelcalibratordoesnotmatchthecalibrator’ssignal(withinspecication),
adjustitasexplainedinthesoundlevelmeter’soperatinginstructions.
Slidetherangeswitchtothe“OFF”positionandremovethemicrophone.
Battery replacement
Replacethebatterywhenthereisnosoundpressureoutputfromthe
soundlevelcalibrator.
Openthebatterycovertoaccessthe9Vbattery.Removethebatteryand
replacewithnew9Vbattery.
ForserviceonthisoranyotherREEDproductorinformationonother
REEDproducts,contactREEDInstrumentsatinfo@reedinstruments.com