Table 2: Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the devices should assure that it is used in
such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC
61000-4-4
± 2 kV for power
supply
lines
± 1 kV for input/
output
lines
± 2 kV for power
supply
lines
NA
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC
61000-4-5
2 kV
0, 90, 180
& 270
degree phase
angles
2 kV
0, 90, 180
& 270
degree phase
angles
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
100% drop, 0.5
periods
(0°, 45°, 90°, 135°,
180°,
225°, 270°, 315°)
100% dip,
1 period
30% dip, 25/30
periods
100% drop, 5
seconds
100% drop, 0.5
periods
(0°, 45°, 90°,
135°, 180°,
225°, 270°, 315°)
100% dip,
1 period
30% dip, 25/30
periods
100% drop, 5
seconds
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the device
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC
61000-4-8
3 A/m 3 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial
or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.