1. Connect the Handpiece to the Handpiece connector socket on the System.
2.
Connect the 64 pin Tip(8mm*8mm) to the handpiece. The 64mm tip is a-sterilized and
single use only.
3.
Before turning the System “ON”, verify that the power cord is connected to the inlet located
on the rear panel of the system and to the main power outlet. Switch on the system by
turning the main power switch located in the rear panel of the system to its “ON” position.
4.
The System loads the software and enters a self-test mode. If any problem is detected
during the test the error message will pop up.
5.
Verify on the screen that the Software version is properly displayed and the connected
Handpiece type is recognized correctly.
6.
The Home screen Fig (1) appears on the screen, when you start the machine. Press Start
on the screen to switch to treatment mode.
7.
Select the treatment parameters.
8.
Press the Ready button to start the treatment.
9.
Apply Handpiece on the area to be treated, ensuring a full contact with pressure.