002
CE Call
Performs the CE Call at the start or end of the service.
[0 or 1 / 0 / 1 /step]
0: Start of the service
1: End of the service
NOTE: This SP is activated only when SP 5816-001 is set to “2”.
003
Function Flag
Enables or disables the remote service function.
[0 to 1 / 0 / 1 /step]
0: Disabled, 1: Enabled
004
Communication Test Call
[- / - / -](Execute)
005
Device Information Call
[- / - / -](Execute)
007
SSL Disable
Controls if RCG (Remote Communication Gate) confirmation is done by SSL during an
RCG send for the @Remote over a network interface.
[0 or 1 / 0 / 1 /step]
0: Yes. SSL not used.
1: No. SSL used.
008
RCG Connect Timeout
Sets the length of time (seconds) for the time-out when the RCG (Remote Communication
Gate) connects during a call via the @Remote network.
[1 to 90 / 30 / 1 second /step
009
RCG Write Timeout
Sets the length of time (seconds) for the time-out when sent data is written to the RCG
during a call over the @Remote network.
[1 to 100 / 60 / 1 second /step]
3. SP Mode Tables
294