EasyManua.ls Logo

Rigol DG812 - Chapter 1 Test Overview; Test Preparations; Recommended Test Devices

Rigol DG812
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Test Overview RIGOL
DG800 Performance Verification Guide
1-1
Chapter 1 Test Overview
Test Preparations
Before performing the test, make sure that the instrument is within the calibration
period (the recommended calibration period is 1 year) and has been warmed up for
at least 30 minutes under the specified operation temperature (18 to 28).
Recommended Test Devices
It is recommended that you use the test devices listed in the table below or other test
devices whose performance specifications satisfy the Performance Requirement
listed in the table below to test the performance specifications of the DG800 series.
Table 1-1 Recommended test devices
Device Performance Requirement
Recommended
Instrument
Frequency Counter
>10MHz
Accuracy: 0.1ppm
Agilent 53131A
6
1
/
2
digits
RIGOL DM3068
Power Meter
-30dBm to +20dBm
Accuracy: ±0.02dB
Resolution: 0.01dB
Agilent E4418B
Spectrum Analyzer
Minimum resolution bandwidth is
100Hz
RIGOL DSA815
Oscilloscope
Bandwidth: 500MHz
Rise/Fall time measurement
function
Overshoot measurement
function
RIGOL DS4000 series
BNC (m)-BNC (m)
--
BNC (m)-Dual banana plug (m)
--
50Ω/1W
--
-30dBm to +20dBm
Agilent N8482A
Used to connect the power meter
and power sensor
--
N (f)-BNC (m)
--
BNC (f)-N (m)
--

Other manuals for Rigol DG812

Related product manuals