Appendix RIGOL
DG800 Performance Verification Guide
7
Spurious Signal Test
Output
Frequency
Start
Frequency
Stop
Frequency
A A-0dBm Specification Pass/Fail
5MHz 0Hz 30MHz <-60dBc
10MHz 0Hz 50MHz <-60dBc
20MHz
[1]
0Hz 100MHz <-54dBc
30MHz
[2]
0Hz 150MHz <-48dBc
Note
[1]
: Only applicable to DG822, DG821, DG832 and DG831.
Note
[2]
: Only applicable to DG832 and DG831.
Rise/Fall Time Test
Waveform Setting Measurement Value Specification Pass/Fail
Square
Frequency: 1kHz
Amplitude: 1Vpp
Offset: 0V
dc
Rise Time
Typical (1 Vpp,
1 kHz)
9 ns
Fall Time
Overshoot Test
Waveform Setting Measurement Value Specification Pass/Fail
Square
Frequency: 100kHz
Amplitude: 1Vpp
Offset: 0V
dc
Typical
(100kHz, 1Vpp)
5%