Technical Data
Influence Quantity Range of Influence Meas ur ing Ranges Attenuation
Co mmon Mo de i n ter fer en ce
vo ltag e
No r mal Mo de
in ter fer en ce r ati o
Noise quantity max. 1000 V dc V dc > 120 dB
Noise quantity max. 1000 V ~
50-60 HZ sinusoidal
6.0 V~,60 V~ >80 dB
600 V~ > 70 dB
1000 V~ > 60 dB
Noise quantity V ~
Value of the measuring range at a
time
Max. 1000V~ ,50Hz, 60Hz
Sinusoidal
V dc > 50dB
Noise quantity max. 1000 V dc V~ >110dB
RISHABH 6012-6016 MANUAL
EMC Immunity IEC 61326-1:2012, Table A.1
Immunity IEC 61000-4-2 : 8 KV atmosphere discharge, 4 KV contact discharge
IEC 61000-4-3 : 3 V/m
Safety IEC 61010-1-2010
IP for water & dust IEC 60529
Pollution degree 2
Installation category
1000 V CATIII / 600 V CATIV, 600V CATII for DMM 6013
High Voltage Test
7.4 kV (IEC 61010-1-2010), 3.5kV For DMM 6013
Test & Procedure IS 13875
Operating Temperature
Storage Temperature
Relative Humidity
IP
Altitude
-10 to +50 C
o
-20 to +70 C
o
<75% non condensing
IP 50 for Housing, IP 20 for Terminal
Up to 2000m