08
1.9 Excerpts from EMC-related guidelines and statements
in manufacturer's accompanying documents
Phenomenon
RF emissions
Harmonic distortion
Voltage fluctuations and flicker
Compliance
CISPR 11 Group 1,Class B
IEC61000-3-2 Class A
IEC61000-3-3 Compliance
Electromagnetic environment
Home healthcare environment
Home healthcare environment
Home healthcare environment
EMS Compliance Table (Table2-5)
EMI Compliance Table(Table 1)
Table 1 - Emission
Phenomenon
Electrostatic
Discharge
Radiated RF
EM field
Proximity fields
From RD
Wireless
communications
equipment
Rated power
Frequency
Magnetic fields
Basic EMC standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Immunity test levels
Home healthcare environment
±8 kV contact
±2 kV,±4 kV,±8 kV,±15 kV air
20 V/m
26 MHz-2.5 GHz
80 % AM at 1 kHz
10 V/m
80 MHz-2.7 GHz
80 % AM at 1 kHz
Refer to table 3
30 A/m
50 Hz or 60 Hz
Table 2 - Enclosure Port
1.Safety precautions