A
161
Table 205: Recommended separation distances (life-supporting devices)
Recommended separation distances between portable and mobile RF communications equipment and the Accu-Chek
Insight insulin pump.
The Accu-Chek Insight insulin pump is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Accu-Chek Insight insulin pump can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Accu-Chek
Insight insulin pump as recommended below, according to the maximum output power of the communication equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2
P
80 MHz to 800 MHz
d = 1.2
P
800 MHz to 2.3 GHz
db = 2.3
P
2.3 GHz to 2.5 GHz
da = 11.7
P
0.01 12 cm 12 cm 23 cm 118 cm
0.1 38 cm 38 cm 73 cm 373 cm
1 1.2 m 1.2 m 2.3 m 11.8 m
10 3.8 m 3.8 m 7.3 m 37.3 m
100 12 m 12 m 23 m 118 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meter (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.