EasyManua.ls Logo

Roche ACCU-Chek Guide Me - Page 16

Roche ACCU-Chek Guide Me
37 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Control Tests
3
4
Select the control
solution to test. You
will confirm the level
later in the test.
5
Remove the bottle
cap. Wipe the tip of
the bottle with a
tissue. Squeeze the
bottle until a tiny
drop forms at the tip.
6
Touch the drop to the
yellow edge of the
test strip. Do not put
control solution on top
of the test strip.
7
A flashing appears
when there is enough
control solution in the
test strip.
8
or
mmol/L
or
mmol/L
The control result, the bottle symbol,
and the flashing L1 or L2 appear on the
display. Press
to confirm the control
level you tested.
Press
or
to alternate between L1
and L2.
9
or
OK appears if the control result is within
range.
Err appears if the control result is out of
range.
10
Wipe the tip of the
bottle with a tissue.
Cap the bottle tightly.
Remove and discard
the used test strip.
81383_09241493001_09241493001_01.indd 1581383_09241493001_09241493001_01.indd 15 10/13/2020 2:44:38 PM10/13/2020 2:44:38 PM

Other manuals for Roche ACCU-Chek Guide Me

Related product manuals