EasyManua.ls Logo

Roche ACCU-CHEK Inform II - Meter reset

Roche ACCU-CHEK Inform II
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting • 11
151
Meter reset A meter reset should only be performed if all other reme-
dies have failed.
1 Place the meter on a level surface.
2 Press the On/Off button for at least 12 seconds.
The meter powers off and on again.
The Roche logo is displayed.
If the Roche logo does not appear within 60 seconds,
place the meter in a base unit for a minimum of
15 minutes to recharge the battery.
The meter performs a system check. During the
system check, the Versions screen appears for a few
seconds, then disappears. The Versions screen
shows the version numbers of the meter compo-
nents. (The Versions screen shown here is for illus-
tration purposes only. Version numbers on your
meter may differ.).
A message appears informing you that the date and
time settings have been lost due to the meter reset.
3 Confirm the message Loss of Date/Time.
Kernel:
RootFS:
App:
Boot:
Build: 04.00.00
Versions
04.00.00
02.05.00
02.05.00
02.05.00
Checking system...
Performing self-checks...
Performing self-checks...
W-1302: D/T settings lost
due to meter reset or loss of
power. Enter D/T manually,
then synchronize meter or
see System Administrator
before performing further
tests.
Loss of Date/Time

Table of Contents