Patient Glucose Testing • 3
53
The following messages
1
may appear in instead of a
numeric test result:
CR LO (below the Critical Range threshold, but
within the Reportable Range)
CR HI (above the Critical Range threshold, but
within the Reportable Range)
RR LO (below the Reportable Range threshold, but
within the System Measurement Range)
RR HI (above the Reportable Range threshold, but
within the System Measurement Range)
LO (below the System Measurement Range)
HI (above the System Measurement Range)
Instructions on how to add comments are provided in the
following section.
1. Results that are below or above the Critical Range thresh-
old, but within the Reportable Range can be configured to
display as either numeric results or as CR LO or CR HI. All
other results exceeding specified limit values will always
be displayed as RR LO, RR HI, LO, or HI.
For test results that lie outside the critical or reporta-
ble range, a message (up to 100 characters in length)
can be configured during setup. This message is then
displayed with the corresponding test results.