EasyManua.ls Logo

Roche ACCU-CHEK Inform II - Confirming or Selecting the Lot Number for Linearity Test Kits; Confirming or Selecting the Test Strip Lot

Roche ACCU-CHEK Inform II
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Linearity Testing • 7
93
Confirming or selecting the lot
number for linearity test kits
You are now prompted to confirm or enter the lot number
of the linearity test kit. Compare the number displayed by
the meter to the number on the label of the linearity test
kit.
3 If you want to use the preselected number dis-
played by the meter, touch to confirm.
To use a different number than the lot number dis-
played, touch to open the keypad and enter
the number manually (see page 85).
Confirming or selecting the test
strip lot
Once you have entered and confirmed the lot number of
the linearity test kit, you are asked to choose the lot
number of the test strips. Compare the number displayed
by the meter to the number on the label of the test strip
vial.
4 If you want to use the preselected number dis-
played by the meter, touch to confirm.
To use a different number than the lot number dis-
played, touch to select the number from a list,
enter the number manually via keypad, or use the
barcode scanner to enter the lot number (see
page 73).
Linearity Test
12345678?
Use Linearity L
ot
12:48 pm
09/18/13
Linearity Test
Linearity 12345678
Use Strip L
ot
12:48 pm
09/18/13
123456?