9
Blood Glucose Tests
w WARNING
Blood glucose results can be displayed
in either mmol/L or mg/dL. The back
label of the meter shows the unit of
measurement. If the meter shows the
wrong unit, contact the Accu‑Chek
Customer Careline on 0800 701 000
(UK) or 1 800 709 600 (ROI). If you do
not know which unit of measurement is
correct for you, contact your healthcare
professional. Using the wrong unit of
measurement may cause
misinterpretation of your actual blood
glucose level and may lead to improper
therapy.
mmol/L or mg/dL is printed here
Using the Accu-Chek Instant
System
w WARNING
• If you drop the meter or drop the
meter with a test strip inserted, the
meter and/or test strip could be
damaged. Discard the test strip and
perform a control test with control
solution and a new, unused test strip
to ensure the meter and test strips
are working properly. Then repeat
the blood glucose test with a new
test strip.
• Do not store test strips in high heat
and moisture areas (bathroom or
kitchen)! Heat and moisture can
damage test strips.
• Use only Accu‑Chek Instant test strips.
• Use the test strip immediately after
removing it from the test strip container.
• Do not apply blood or control solution to
the test strip before inserting it into the
meter.
• Close the test strip container tightly
immediately after removing a test strip
to protect the test strips from humidity.
• Store the unused test strips in their
original test strip container with the cap
closed.
• Check the use by date on the test strip
container. Do not use the test strips after
that date.
• Store the test strip container and meter
in a cool, dry place such as a bedroom.
• Refer to the test strip package insert for
test strip storage and system operating
conditions.
82688_09291423001_01.indd 982688_09291423001_01.indd 9 8/20/2020 1:58:03 PM8/20/2020 1:58:03 PM