cobas p 512 Part B - 0BSystem description
Operator's Manual - Version 1.6 - 10/2015 2-85
Figure 2-61: Error history menu Error listing
The error message selected in the picture above provides the following information:
Meaning
Description
F
Fatal error
164840
System time when the error occurred
1211
Date when the error occurred
stat_mgr0
Name of the program that issued the error
0122
Error message number
03
Internal error number QNX
0000
Internal error number Roche PVT
122 Task <comm_astm0> not started
Error description
NOTICE
The error log file information can be helpful to trace errors during installation and system setup
and for general troubleshooting. Refer to the Troubleshooting chapter for detailed information
about error messages and remedies.
[Reset] button
The current error list will be cleared after pressing the [Reset] button.