cobas p 512 Part D - Maintenance
Operator's Manual - Version 1.6 - 09/2015 4-25
Figure 4-29: Removing the mains supply cable
6. Using a thin-blade screwdriver, push up the tab (on the lower part of the fuse holder)
and pull out the fuse holder.
Figure
4-30: Push up the tab ...
Figure
4-31: and remove the fuse holder
7. Replace the fuse. Use only fuses that comply with the following specifications:
230 Volt power supply: T 10 AH 250 V (size: 5*20 mm)
115 Volt power supply: T 10 AH 250 V.
8. Push the fuse holder into the housing until you hear an audible click.
Figure 4-32: Inserting the fuse holder until an audible click can be heard
9. Plug the connector of the mains supply cable into the mains supply connection on the
system. That concludes the procedure for replacing the fuse.