cobas p 512 Part B - 0BSystem description
Operator's Manual - Version 1.6 - 10/2015 2-59
Racks and workplace areas on the drawer
Depending on the customer configuration, one or multiple workplaces can be assigned for a
single drawer. A workplace can be a part of a single rack, a single rack or multiple racks.
Thus, you can see distinct sections of the drawer shown in the software when multiple
workplaces have been setup.
Each rack is indicated by a number or letter. Thus a sample can be located exactly if the
patient data is later searched.
Figure
2-46: Drawer with four separate
racks/workplaces
Figure
2-47: Drawer with one rack/workplace
The rack symbols also have the function of a button. After pushing in the drawer in the
output sorter, it cannot be manually pulled out again. By pressing on the rack symbol, the
drawer can be ejected and the processing information can be deleted.
If the process has been activated by mistake, it can be cancelled by pressing on the picture of
the corresponding rack again. The opened drawer will then have to be closed again in order
for the sorting process to continue at the next available position within the rack.