MagNA Pure 96 System, Customer Training Guide V1.0
32
Setting up a Run for Purification of Genomic DNA from Whole Blood Samples
B
B
Defining the Parameters in the Software
1.2 Defining the Parameters in the Software
Before starting a purification run, you must provide the software with the necessary information on pro-
tocol and sample parameters. The most important and obligatory information is:
Order type
Kit name
Purification protocol
Sample and elution volume
Number and name of samples
Once it has this information, the software calculates the type and amount of isolation reagents and dis-
posables.
To create the order
In the Overview tab click . The Orders sub-tab opens.
Above the Test Parameter area, select Purification from the order type list.
Select the DNA/Viral NA SV kit from the MagNA Pure Kit Name list. The Protocol list is displayed.
Select the DNA Blood SV protocol from the Protocol list. Additional options are displayed.
In the Volume group, select 200 in the Sample list.
In the Volume group, select 50 in the Elution list.
Select IC3 from the Internal Control list.
Select None from the Custom Source Plate list.
Select MP96 Output Plate from the Target Plate list.