Maintenance and Trouble Shooting
87
C
C
4.2 Instrument Error Codes
Valid for Software Version 1.1 and higher
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation Action
1 to 99 [diverse
messages]
- Internal error Shut down instrument
at main switch and
restart system. If problem
persists, contact Technical
Service.
3502 Barcode ID x
is invalid
The entered barcode ID is
not valid.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
The entered barcode ID is
not valid.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
3504 Liquid Waste
Discard
aborted.
Liquid Waste Discard run
was aborted.
Liquid Waste Discard run
was aborted by Operator.
Restart Liquid Waste
Discard.
3505 Front door is
open.
Front door is not properly
closed. Check front door.
Front door is not properly
closed. Check front door.
Check front door. Open
and close dront door again.
If problem persists, contact
Technical Service.
3506 Invalid kit
code.
The entered kit name coded
in the cartridge ID barcode
is invalid.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
The entered kit name
coded in the cartridge
ID barcode is invalid.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
3507 Invalid
cartridge
expiry date
format.
The entered cartridge expiry
date coded in the cartridge
ID barcode is invalid.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
The entered cartridge
expiry date coded in the
cartridge ID barcode is
invalid.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
3508 Reagent
Cartridge
expired.
The cartridge expiry date
coded in the cartridge ID
barcode is exceeded.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
The cartridge expiry date
coded in the cartridge ID
barcode is exceeded.
It is recommended to
discard the cartridge. If you
decide to use the expired
cartridge the appropriate
channel will be flagged
FAIL on the result screen.
Note:
the function of
expired cartridges cannot
be guaranteed.
Error Codes and Trouble Shooting Guide
Instrument Error Codes