MagNA Pure Compact Operator’s Manual - Version 1.3
88
C
C
3509 Different kit
types mixed.
Mixing up of different kit
types is not possible.
Mixing up of different kit
types is not possible.
Make sure that only one kit
type is used per run.
Repeat entering of the
barcode ID, or type in
barcode manually. If
problem persists, contact
Technical Service.
3510 Too many data
found.
The search found too many
data.
Search more specific.
More than 1.000 data sets
were found.
Search more specific.
3511 Unable to save
to folder.
Unable to save to folder.
Check folder name in set-
tings.
Unable to save to folder. Check folder name in
settings. If problem
persists, contact
Technical Service.
3512 Front door is
open.
Front door is open and
should be closed.
Front door should be closed
during initialization of
instrument and during
purification run.
Close front door and
press OK.
3513 Front door
cannot be
lokked.
Front door cannot be
locked.
Failure of front door lock or
door lock sensor.
Open and close front door
again. If problem persists,
contact Technical Service.
3514 Z crash sensor
test failed.
Z crash sensor
malfunction.
Z crash sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3515 Pressure sen-
sor test failed.
Pressure sensor
malfunction.
Pressure sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3516 Magnet home
position sen-
sor failed.
Magnet home position
sensor malfunction.
Magnet home position
sensor is not working
properly.
Restart system. If problem
persists, contact Technical
Service.
3517 Magnet ON
sensor failed.
Magnet ON sensor
malfunction.
Magnet ON sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3518 Z axis home
position
sensor failed.
Z axis home position sensor
malfunction.
Z axis home position sensor
is not working properly.
Restart system. If problem
persists, contact Technical
Service.
3519 Y axis limit
sensor failed.
y axis limit sensor
malfunction.
Y axis limit sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
3520 Y axis home
position
sensor failed.
Y axis home position sensor
malfunction.
Y axis home position sensor
is not working properly.
Restart system. If problem
persists, contact Technical
Service.
3521 P axis home
position
sensor failed.
P axis home position sensor
malfunction.
P axis home position sensor
is not working properly.
Restart system. If problem
persists, contact Technical
Service.
3522 Waste tank
sensor failed.
Waste tank sensor
malfunction.
Waste tank sensor is not
working properly.
Restart system. If problem
persists, contact Technical
Service.
Error
Code
ID
Error
Message
Error Information
(displayed on
information screen)
Explanation Action
Error Codes and Trouble Shooting Guide
Instrument Error Codes
MagNA Pure Compact Manual.book Seite 88 Dienstag, 28. Februar 2006 4:16 16