5
Important Information regarding the MagNA Pure Compact Instrument Operator’s Manual
Chapter A:
2. Installation of the Instrument
2.4 Setup, page 16
Current Version Changes
2.4.1
Selecting a
Location
The MagNA Pure Compact Instrument
requires very little setup. Choose for
the MagNA Pure Compact Instrument
a clean, dry, level, stable surface within
3 m of a compatible electrical outlet.
To ensure proper ventilation, leave 10
cm of space behind the instrument and
15 cm at each side of the instrument.
No space is needed at the back of the
instrument.
The MagNA Pure Compact Instrument
requires very little setup. Choose for
the MagNA Pure Compact Instrument
a clean, dry, level, stable surface within
3 m of a compatible electrical outlet. To
ensure proper ventilation, leave 10 cm
of space behind the instrument and
15 cm at each side of the instrument.
3. Description of the Instrument
3.3.5 MagNA Pure Compact Accessories
7 Barcode Scanner – Scanner Specifi cations, page 33
Current Version Changes
Drop Resistance IEC 68-2-32 Test ED; withstands
repeated drops from 1.8 m onto a
concrete surface
DIN EN 60068-2-32 Test ED;
withstands repeated drops from 1.8 m
onto a concrete surface