General Overview
15
A
A
2.3.2 Maintenance Items
Note:
To ensure that they work correctly, you must grease the O-rings once per week (or
after every 20 purification runs). If the O-rings malfunction, the instrument nozzles
may leak. For details, please see the Maintenance section in Chapter C.
Perform a leakage test every two months (or after every 50 purification runs) to deter-
mine whether the O-rings should be changed. For details please see the Maintenance
section in Chapter C.
There is an O-Ring Exchange Tool available, which simplifies the O-ring changing pro-
cess. Please contact your local Roche Diagnostics representative if you need O-rings
changed or want one of these tools.
Elution Tubes (35 tubes per package, 2.0 ml each, with stand, with barcode)
Caps for Elution Tubes (35 caps per caps)
Tip Trays (including Piercing Tool and tips). The tips are placed in the following order
into the Tip Tray: one Piercing Tool, one small tip, two large tips.
Cat. No. Purification Kit Contents
03 561 429 001 MagNA Pure LC O-Ring Maintenance Kit O-rings (12 x 8),
High Vacuum Grease (1)
03 753 166 001 MagNA Pure Compact Tip Tray Kit 10 Tip Trays
Kit content:
Installation of the Instrument
Additional Required Items for Operation and Maintenance
MagNA Pure Compact Manual.book Seite 15 Dienstag, 28. Februar 2006 4:16 16