B
Table 1 below shows the systems examined during the self-test. If any test is not passed,
the screen will display an error message. (For an overview of error messages, please see
chapter C.)
Table 1: Overview of self-test
Note: If the self-test is not passed and the remedy as described in the section about errors
and troubleshooting does not help, please contact your local Roche representative.
Note: Besides the start up self-test, the instrument
will automatically perform a self-test after a certain
period of instrument inactivity. You must confirm
execution of this self-test on a message window.
Function Test
Instrument communication Check for correct communication between hardware and
software.
Door sensor Function test of sensor for locked front door.
Magnetic sensor ON/OFF status must correspond to its position
Home sensor Function test of sensor for robotic arm in home position.
Waste tank sensor Check for presence of waste tank.
Housing fan sensor Check for presence of housing fan.
Handling the MagNA Pure Compact Software
Starting the MagNA Pure Compact Instrument