How to Operate the MagNA Pure Compact Instrument
41
B
B
Table 3: Overview of Instrument Statuses
Table 4: Overview of menu and submenu structure
State Description, instrument action
Ready Self-tests OK, instrument is ready for operation, maintenance,
or service functions.
Error Instrument is shut down by fatal error or awaiting user intervention
after error
Running Protocol is beeing processed
Run completed Run is completed, result screen is displayed.
Leaky If the last leakage test has not been successful (state remains until
next O-ring change or next successful leakage test). Performing a
purification run is not possible.
Submenu Function Described in Detailed Actions
Run Performing a purification run Chapter B Perform the sequence of steps on
the Sample Ordering Screens:
䉴 Operator Authorization
䉴 Sample Ordering 1
䉴 Sample Ordering 2
䉴 Sample Ordering 3
䉴 Sample Ordering 4 [available with IC]
䉴 Sample Ordering 5
䉴 Confirmation
Setup Setting basic information Chapter A Setting:
䉴 Time and Date
䉴 Operator administration
䉴 Laboratory name
䉴 Folder where results are saved
䉴 Sample material
Documentation Organizing
documentation
Chapter B Documentation via:
䉴 Printouts
䉴 Saved result files
Maintenance Performing instrument
maintenance
Chapter C Maintaining:
䉴 Leakage Test
䉴 Counter and Reminder
䉴 Error Log
䉴 O-ring Exchange
䉴 Liquid Waste Discard
䉴 UV Decontamination
Service Performing specialized service
functions
Not described in this
Operator’s Manual.
Used by Roche service operators only
Handling the MagNA Pure Compact Software
Overview on Main Menu Screen
MagNA Pure Compact Manual.book Seite 41 Dienstag, 28. Februar 2006 4:16 16