MagNA Pure Compact Operator’s Manual - Version 1.3
54
B
B
How to Operate the MagNA Pure Compact Instrument
B
B
Do one of the following:
SAMPLE ORDERING SCREEN 3 appears. While at
this screen, do the following:
Pipet the correct type and amount of sample
material (as chosen on Sample Ordering Screen
2) into a sample tube (tube without barcode label,
from the reagent kit).
Identify the tube by attaching your own barcode
label or writing a number on the tube with a per-
manent marker.
Put the filled sample tube in position 1 on row 1
of the Tube Rack.
Note: It is recommended to use the
appropriate row in the tube rack for original
sample tubes, if suitable for these tubes
(e.g. max. 15 ml Falcon tubes).
On SAMPLE ORDERING SCREEN 3, touch Sample ID field 1. In that field, enter the ID of Sample 1 (i.e., the one
you just put in the rack) by either scanning the (self-attached) barcode or using the virtual keyboard to type the
ID number. (Touch the Keyboard button to access the keyboard.)
(optional) In the Comment field, you may enter
additional information about Sample 1. (Touch the
field to highlight it, then use the virtual keyboard.)
Repeat the above sample placement and
identification steps for each sample to be purified
(e.g., for Sample position 2, 3, …, 8).
Result: As you fill in the "Sample-ID" field for each
sample, the corresponding position in the reagent
layout will change from green (active) to dark blue
(installed in instrument).
Action Handling Software programming
9
If ... Then ...
You are not using internal
controls (i.e., you selected
a protocol with no internal
control on SAMPLE
ORDERING SCREEN 2)
(Action 8 above)
Insert the tube rack into the MagNA Pure
Compact Instrument.
Touch the "Samples inserted" confirmation
button (a).
The Next button (b) will now be active. Touch
this button to confirm the information on this
screen and go to the next programming
screen (SAMPLE ORDERING SCREEN 5).
Go to Action 11 below.
You are using internal
controls (i.e., you selected
a protocol with internal
control on SAMPLE
ORDERING SCREEN 2)
Touch the "Sample inserted" confirmation
button (a).
The Next button (b) will now be active. Touch
this button to confirm the information on this
screen and go to the next programming
screen (SAMPLE ORDERING SCREEN 4).
Go to Action 10 below.
Performing a Purification Run
Purification
Performing a Purification Run
Purification
55
4_MagNA Pure Compact_ChapterB_2.fm Seite 54 Mittwoch, 8. März 2006 10:13 10