E-16 Operators Manual · Version 1.0
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
27 Sample Barcode
Sample Barcode Label
Error
1-5 WARNING 1. Check whether the barcode label is
dirty.
2. Check whether the barcode is
positioned correctly.
3. Check whether the barcode label is
missing.
4. Check whether the barcode label has
the correct type of code.
5. Contact Technical Support.
28 Barcode Reader
Comm.
Barcode Reader Error
1-30 S.STOP 1. Check whether the barcode label has
the correct type of code.
2. Contact Technical Support.
29 Reflectance ADC
ADC Reflectance Error
1 WARNING 1. Contact Technical Support.
2-7 STOP 1. Contact Technical Support.
30 SG ADC
ADC Specific Gravity
Error
1 WARNING 1. Contact Technical Support.
2-7 STOP 1. Contact Technical Support.
31 CLA ADC
ADC Clarity Error
1 WARNING 1. Contact Technical Support.
2-7 STOP 1. Contact Technical Support.
33 Wavelength Setting?
Wavelength Error
2-10, 12 WARNING 1. Correctly certify the wavelength and
correction value with the blank pad.
2. The subcategory codes represent the
items listed as follows:
Subcategory Item
2pH
3LEU
4NIT
5PRO
6GLU
7KET
8UBG
9BIL
10ERY
11COL
12CLA
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)