Roche Diagnostics
Operators Manual · Version 1.0 C-11
URISYS 2400 5 Daily operation
Termination of analysis
appears (Warning/Data overwrite). The analyzer changes to S.Stop or
Stand-By status.
Delete the results of samples with sequence numbers higher than the actual
sequence number.
Otherwise, analysis of samples cannot continue.
Termination of analysis
Completion of analysis
After measuring the last test strip, the analyzer changes to Stand-By.
To continue sampling, place more racks with sample tubes on the input line,
and press <Start> to begin analysis again. The analyzer continues sampling
and measurement. For more information on the processing of additional
samples, see:
Interrupting analysis
Using the S. Stop button on the [Routine Monitor] screen, the operator can
interrupt analysis. Racks in the cross-feed line are still processed; then the
analyzer turns to Sampling Stop status. Racks in the input line remain in their
current position. Samples that have already been pipetted onto test strips are
measured.
a To interrupt analysis
1 Press <S.Stop> on the [Routine Monitor] screen. The confirmation
window appears.
2 Press <OK> to stop sampling.
To enter the routine sequence number, STAT sequence number and operator ID on
page C-6
Delete routine or STAT sample data on page C-27
Do not open the cassette compartment door or the strip waste container door
during analysis. Otherwise, the analyzer stops and moves to Stand-By status. When
<Start> is pressed again, the subsequent reset clears the racks still present in the
cross-feed line without sampling them, and removes test strips still present on the
TS transfer base without measuring them.
Process additional samples on page C-13