R&S
®
CMW 500 Remote Control
Control of the Instrument
Operating Manual 1202.3986.32 – 03 84
Measurement control commands are of the following type (see also Firmware
Applications):
SCPI subsystem <Application> Measurement instance Context
INITiate
ABORt
STOP
:GSM |
:WCDMa | :GPRF
:MEASurement<i> :POWer
:EPSensor
:MEValuation
...
(depending on <Application>)
Example: INITiate:GPRF:MEASurement:POWer
5.4.1.2 INITiate:<Application>:MEASurement<i>
Starts the measurement in the repetition mode set via
CONFigure:<Application>:MEASurement<i>:<Context>:REPetition
(single shot or continuous, see Statistical Settings). The command resets the counters
for the evaluation period and statistics cycle to zero.
INITiate... can be used irrespective of the current measurement state. An initiated
measurement reserves all necessary system resources and enters the "RUN" state. If
the hardware and system resources are already assigned to another firmware
application, this firmware application is released in order to start the measurement; see
Resource and Path Management.