R&S
®
CMW 500 Command Reference
Instrument-Control Commands
Operating Manual 1202.3986.32 – 03 130
SYSTem:KLOCk <Enable>
Locks or unlocks the local controls of the instrument. This includes the front panel
keys, the keyboard, or other local interfaces.
<Enable> ON: Local key locked (key lock enabled)
OFF: Local keys unlocked
Def. value: OFF
Example SYSTem:KLOCk ON
Lock the local keys.
Characteristics Firmware version V1.00
SCPI-confirmed
SYSTem:PRESet
Sets the instrument parameters to values for good local/manual interaction. The
command corresponds to the Preset button in the Reset dialog.
Example SYSTem:PRESet
Force the R&S CMW 500 to a reset state that is optimized for
manual operation.
*RST
Optimize the R&S CMW 500 for remote operation.
Characteristics Firmware version V1.00
SCPI-confirmed, no query
SYSTem:TIME <Hour>, <Minute>, <Second>
Sets the time of the internal clock.
<Hour>
<Minute>
<Second>
Hour, 0 to 23
Minute, 0 to 59
Second, 0 to 59 (60 due to rounding)
Numbers entered are rounded to the closest integer value.
Def. value: – (a *RST does not affect the time setting)
Example SYSTem:TIME?
Query the current time. Possible response: 15, 09, 20
Characteristics Firmware version V1.00
SCPI-confirmed
SYSTem:VERSion?
Returns the SCPI version number to which the instrument complies. The instrument
complies to the final SCPI version 1999.0.