CMW 500 GPRF Applications
GPRF GUI Reference
Operating Manual 1202.3986.32 – 03 149
List Configuration > Current Index
Number of frequency/level step in the sequence list. This parameter is available in
"Step" mode only.
GPIB command:
SOURce:GPRF:GENerator<i>:LIST:CINDex
List Configuration > Mode
Defines how the RF generator runs through the list (or through the list section between
the "Start Index" and the "Stop Index").
Auto The generator cycles through the list section between the "Start
Index" and the "Stop Index", transmitting at each
frequency/level step for the selected "Dwell Time". The cycle is re-
started automatically from the beginning of the list.
Single The generator runs through the list section
between the "Start Index"
and the "Stop Index" for a single time. Transmission at the start
frequency/level step must be initiated using the "Execute Single"
hotkey.
Step Manual selection of a single frequency/level step. The RF generator
transmits at the frequency/level with the "Current Index" until a new
current index is selected. The dwell time is ignored.
List Mode Settings