R&S
®
CMW 500 GPRF Applications
GPRF GUI Reference
Operating Manual 1202.3986.32 – 03 155
CMW 500 provides the following filters of fixed shape:
The RRC filter with a roll-off = 0.22 and a 3.84 MHz bandwidth specified for
WCDMA TX tests ("WCDMA" filter, see standard 3GPP 34.121).
The 1.4 MHz-wide baseband receiver filter specified for CDMA 2000 TX tests
("CDMA" filter; see standard TIA/EIA/IS-2000.2-A).
GPIB command:
CONFigure:GPRF:MEASurement<i>:POWer:FILTer:TYPE
CONFigure:GPRF:MEASurement<i>:POWer:FILTer:GAUSs:BWIDth
CONFigure:GPRF:MEASurement<i>:POWer:FILTer:BANDpass:BWIDth
7.2.2.4 Power Configuration: List Configuration
The "List Configuration" parameters activate and configure the list mode of the
GPRF "Power" measurement.
Power Configuration: List Configuration
List Mode
Selects the list mode. "Off" means that the "Power" measurement yields results at a
single power/frequency step. If the "List Mode" is switched on, the R&S CMW 500
measures a list of power steps in accordance with the "List Mode" settings.
GPIB command:
CONFigure:GPRF:MEASurement<i>:POWer:LIST
List Section
Defines the step range to be used in list mode. The R&S CMW 500 measures all steps
between the Start Index and the Stop Index in the List; the number of steps measured
(Count = Stop Index – Start Index + 1) is displayed for information.
GPIB command:
CONFigure:GPRF:MEASurement<i>:POWer:LIST:STARt
CONFigure:GPRF:MEASurement<i>:POWer:LIST:STOP
CONFigure:GPRF:MEASurement<i>:POWer:LIST:COUNt?
Freq. / Exp. Nom. Pow. List
Definition of 500 not necessarily different pairs of frequencies and levels. The value
range for the frequency covers the entire configurable value range of the instrument,
深圳德标仪器 135-1095-0799