Operating Manual 1202.3986.32 – 03 168
After a *RST, the measurement is switched off. Use
READ:GPRF:MEASurement:POWer? to initiate a single-shot measurement and
retrieve the results.
Advanced settings
You can also run the measurement in continuous mode and retrieve the results using
FETCh:GPRF:MEASurement:POWer...? (see Measurement Control). A reliability
indicator is returned with each measurement result.
7.3.2.2 Specifying General Measurement Settings
FPRINT ************************************************************
FPRINT System-Reset
FPRINT ************************************************************
*RST;*OPC?
*CLS; *OPC?
FPRINT ************************************************************
FPRINT Define signal routing, perform RF and analyzer settings
FPRINT for a 20 dBm GSM signal at a GSM900-specific frequency (channel no. 1)
FPRINT ************************************************************
ROUTe:GPRF:MEASurement:RFSettings:CONNector RF1C
CONFigure:GPRF:MEASurement:RFSettings:EATTenuation 2
CONFigure:GPRF:MEASurement:RFSettings:ENPower 20
CONFigure:GPRF:MEASurement:RFSettings:UMARgin 5
CONFigure:GPRF:MEASurement:RFSettings:FREQuency 890.2E+6
7.3.2.3 Specifying Measurement-Specific Settings
FPRINT ************************************************************
FPRINT Define step in accordance with GSM timeslot
FPRINT and measurement length
FPRINT ************************************************************
CONFigure:GPRF:MEASurement:POWer:SLENgth 577.9230769E-6
CONFigure:GPRF:MEASurement:POWer:MLENgth 400E-6
FPRINT ************************************************************
FPRINT Define measurement statistics (5 GSM timeslots)
FPRINT ************************************************************
CONFigure:GPRF:MEASurement:POWer:SCOunt 5
FPRINT ************************************************************
FPRINT Select 30 kHz Gauss measurement filter
FPRINT ************************************************************
CONFigure:GPRF:MEASurement:POWer:FILTer:TYPE GAUSs