R&S
®
CMW 500 GPRF Applications
GPRF Programming
Operating Manual 1202.3986.32 – 03 170
CONFigure:GPRF:MEASurement:POWer:LIST:FREQuency? 0
CONFigure:GPRF:MEASurement:POWer:LIST:FREQuency 0, 7.55E+8
CONFigure:GPRF:MEASurement:POWer:LIST:ENPower? 0
CONFigure:GPRF:MEASurement:POWer:LIST:ENPower 0, -5
7.3.2.6 Single Shot and Continuous Power Measurements
FPRINT ************************************************************
FPRINT Start single-shot measurement, return RMS-averaged power in last measured
timeslot.
FPRINT Return standard deviation of the RMS power results (without repeating the
measurement.
FPRINT Query the measurement state (should be "RDY").
FPRINT ************************************************************
INIT:GPRF:MEASurement:POWer
FETCh:GPRF:MEASurement:POWer:RMS:CURRent?
FETCh:GPRF:MEASurement:POWer:SDEViation?
FETCh:GPRF:MEASurement:POWer:STATe?
FPRINT ************************************************************
FPRINT Start continuous measurement, return last valid result for
FPRINT the RMS-averaged power in last measured timeslot of the current
measurement cycle.
FPRINT Query measurement state and substates (should be "RUN,ADJ,ACT").
FPRINT ************************************************************
CONFigure:GPRF:MEASurement:POWer:REPetition CONTinuous
FETCh:GPRF:MEASurement:POWer:RMS:CURRent?
FETCh:GPRF:MEASurement:POWer:STATe:ALL?
7.3.3 Programming the External Power Sensor Measurement
The following examples show how to configure the External Power Sensor
measurement via a remote-control program and retrieve measurement results.
Key Features
Performing a Single Shot Measurement
Zeroing the External Sensor
Continuous Sensor Measurement
For an introduction to instrument programming refer to the "Remote Control" chapter,
in particular to section Measurement Control.