R&S
®
CMW 500 GPRF Applications
GPRF Command Reference
Operating Manual 1202.3986.32 – 03 187
<Length> Range: 0.010 ms to 1 s
Def. value: 300 µs
Unit: s
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
7.4.3.4 Statistical Settings
The following commands specify the scope of the measurement.
CONFigure:GPRF:MEAS<i>:POWer:REPetition <Mode>
Specifies the repetition mode of the measurement; see Statistical Settings. The
repetition mode specifies whether the measurement is stopped after a single shot or
repeated continuously. Use CONFigure:..:MEAS<i>:...:SCOunt to determine the
number of measurement intervals per single shot.
<Mode> SINGleshot: Single shot measurement
CONTinuous: Continuous measurement
Def. value: SINGleshot
Example See Single Shot and Continuous Power Measurements
Characteristics Firmware version V1.00
CONFigure:GPRF:MEAS<i>:POWer:SCOunt <Count>
Specifies the statistic count of the measurement; see Statistical Settings. The statistic
count is equal to the number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous to
select either single shot or continuous measurements.
<Count> Number of measurement intervals. A measurement interval
comprises a single power/frequency step.
Range: 1 to 1000
Def. value: 10
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
7.4.3.5 Filter Settings
The following commands select the measurement filter (IF filter) for the measurement.