R&S
®
CMW 500 GPRF Applications
GPRF Command Reference
Operating Manual 1202.3986.32 – 03 190
Characteristics Firmware version V1.00
CONFigure:GPRF:MEAS<i>:POWer:TRIGger:TOUT <Trigger Timeout>
Sets a time after which the initiated measurement is automatically terminated if it does
not receive a trigger event.
<Trigger Timeout> Range: 0.010 s to 300.000 s
Def. value: 0.100 s
Def. unit: s
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
CONFigure:GPRF:MEAS<i>:POWer:TRIGger:MODE <Trigger Mode>
Selects the measurement sequence that is triggered by each single trigger event. This
setting is not valid for free run measurements
(CONFigure:GPRF:MEAS<i>:POWer:TRIGger:SOURce FRUN).
<Trigger Mode> ONCE: Trigger once
SWEep: Retrigger Sweep
ALL: Retrigger All
Def. value: ONCE
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
7.4.3.7 List Configuration
The following commands select the sequence mode of the "Power" measurement and
configure the list mode.
CONFigure:GPRF:MEAS<i>:POWer:LIST <Enable>
Enables or disables the list mode for the "Power" measurement.
<Enable> OFF: List mode off (single power step)
ON: List mode on
Def. value: OFF
Example See Configuring the List Mode
Characteristics Firmware version V1.00
CONFigure:GPRF:MEAS<i>:POWer:LIST:COUNt? <Result Count>
Queries the total number of results, depending on the list mode